Inductively Coupled Plasma(ICP) 
                            Atomic/optical emission spectroscopy(ICP-AES/OES), 
                            MASS Spectrometers(ICP-MS). 
                            
                            
                            https://en.wikipedia.org/wiki/Inductively_coupled_plasma | 
                     
                    
                        | 
                             
                                
                                An inductively 
                                coupled plasma (ICP) 
                                or transformer 
                                coupled plasma (TCP)[1] is a type of plasma source in which the energy is supplied by electric currents which are produced by electromagnetic 
                                induction, 
                                that is, by time-varying magnetic fields.[2]. 
                            
                                Plasma electron temperatures can range between ~6 000 K and ~10 000 
                                K (~6 eV - ~100 eV),[4]comparable to 
                                the surface of the sun. ICP discharges are of relatively high electron density, 
                                on the order of 1015 cm−3. As a result, ICP discharges have wide applications where 
                                a high-density plasma (HDP) is needed. 
                            
                                
                                
                                
                                ·        
                                
                                
                                ICP-AES, a type of atomic emission spectroscopy. 
                                ·       
                                
                                
                                ICP-MS, a type of mass spectrometry. 
                                ·       
                                
                                
                                ICP-RIE, a type of reactive-ion etching. 
                                ·       
                                Exciting a beam of noble gas to the metastable state.[5] 
                            
                                Another benefit of ICP discharges is that they are relatively free of 
                                contamination because the electrodes are completely outside the reaction 
                                chamber. By contrast, in a capacitively 
                                coupled plasma (CCP), the 
                                electrodes are often placed inside the reactor and are thus exposed to the 
                                plasma and subsequent reactive chemical species. 
                         | 
                        
                            
                            
                            
                            
                            
                            
                            
                            
                             
                             
                            
                            
                             Picture of an analytical ICP torch 
                             
                             
                            
                             
                             
                            High performance optical system | 
                     
                 
                
                
                
                
             | 
         
     
    
        
            | 
                 
                    Applications : 
                    Examples of the application of ICP-AES include the determination of metals in 
                    wine,[4] arsenic in 
                    food,[5] and trace 
                    elements bound to proteins.[6] 
                
                    ICP-OES is widely used in minerals processing to provide the data on grades of 
                    various streams, for the construction of mass balances. 
                
                    In 2008, the technique was used at Liverpool University to demonstrate that a Chi Rho amulet found in Shepton Mallet and previously believed to be among 
                    the earliest evidence of Christianity in England,[7] only dated to 
                    the nineteenth century.[8][9][10] 
                
                    ICP-AES is often used for analysis of trace elements in soil, and it is for that 
                    reason it is often used in forensics to ascertain the origin of soil samples 
                    found at crime scenes or on victims etc. Taking one sample from a control and 
                    determining the metal composition and taking the sample obtained from evidence 
                    and determine that metal composition allows a comparison to be made. While soil 
                    evidence may not stand alone in court it certainly strengthens other evidence. 
                
                    It is also fast becoming the analytical method of choice for the determination 
                    of nutrient levels in agricultural soils. This information is then used to 
                    calculate the amount of fertiliser required to maximise crop yield and quality. 
                
                    ICP-AES is used for motor oil analysis. Analyzing used motor oil 
                    reveals a great deal about how the engine is operating. Parts that wear in the 
                    engine will deposit traces in the oil which can be detected with ICP-AES. 
                    ICP-AES analysis can help to determine whether parts are failing. In addition, 
                    ICP-AES can determine what amount of certain oil additives remain and therefore 
                    indicate how much service life the oil has remaining. Oil analysis is often used 
                    by fleet manager or automotive enthusiasts who have an interest in finding out 
                    as much about their engine's operation as possible. ICP-AES is also used during 
                    the production of motor oils (and other lubricating oils) for quality control 
                    and compliance with production and industry specifications. 
             | 
            
                
                
                
                
                
                
                
                
                
                
                
                 
                
             | 
         
        
            
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                 
                
                
             | 
         
        
            
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                 
                
                
             | 
         
     
    
        Inductively coupled plasma mass 
        spectrometry 
        
        
        https://en.wikipedia.org/wiki/Inductively_coupled_plasma_mass_spectrometry 
            
                
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                     
                    
                    
                 | 
                
                     
                        Inductively coupled plasma mass spectrometry (ICP-MS) is a type of mass spectrometry which is capable of detecting metals and several non-metals at concentrations as low as one part 
                        in 1015(part per quadrillion, ppq) on non-interfered 
                        low-background isotopes. This is achieved by ionizing the sample with inductively 
                        coupled plasma and then using a mass spectrometer to separate and quantify those ions. 
                    
                        Compared to atomic 
                        absorption spectroscopy, ICP-MS has greater speed, precision, and 
                        sensitivity. However, compared with other types of mass spectrometry, such as thermal 
                        ionization mass spectrometry (TIMS) 
                        and glow discharge mass spectrometry (GD-MS), ICP-MS introduces many interfering 
                        species: argon from the plasma, component gases of air that leak through the 
                        cone orifices, and contamination from glassware and the cones. 
                        
                        The variety of applications exceeds that of inductively coupled plasma atomic emission 
                        spectroscopy and includes isotopic speciation.[citation 
                        needed] Due to possible applications in nuclear technologies, 
                        ICP-MS hardware is a subject for special exporting regulations. 
                    
                 | 
             
         
    
    
        Applications 
            
                
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                     
                    
                    
                 | 
                
                    
                    One of the largest volume uses for ICP-MS is in the medical and forensic field, 
                    specifically, toxicology.[citation 
                    needed] A physician may order a metal assay 
                    for a number of reasons, such as suspicion of heavy metal poisoning, metabolic 
                    concerns, and even hepatological issues. Depending on the specific parameters 
                    unique to each patient's diagnostic plan, samples collected for analysis can 
                    range from whole blood, urine, plasma, serum, to even packed red blood cells. 
                    Another primary use for this instrument lies in the environmental field. Such 
                    applications include water testing for municipalities or private individuals all 
                    the way to soil, water and other material analysis for industrial purposes. In 
                    the forensic field, glass ICP-MS is popular for glass analysis.[citation 
                    needed] Trace elements on glass can be 
                    detected using the LA-ICP-MS. The trace elements from the glass can be used to 
                    match a sample found at the crime scene to a suspect. | 
                
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                    
                     
                    
                    
                 | 
             
         
    
    
        
            
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                 
                
                
             | 
            
                
                In recent years, industrial and biological monitoring has presented another 
                major need for metal analysis via ICP-MS. Individuals working in plants where 
                exposure to metals is likely and unavoidable, such as a battery factory, are 
                required by their employer to have their blood or urine analyzed for metal 
                toxicity on a regular basis. This monitoring has become a mandatory practice 
                implemented by OSHA, 
                in an effort to protect workers from their work environment and ensure proper 
                rotation of work duties (i.e. rotating employees from a high exposure position 
                to a low exposure position). 
                
                11 Signs You Have Heavy Metal Toxicity & How To Prevent It :
                 
                
                
                http://www.naturallivingideas.com/heavy-metal-toxicity-signs/ 
                
             | 
         
     
    
        
            | 
                
                ICP-MS is also used widely in the geochemistry field for radiometric dating, in 
                which it is used to analyze relative abundance of different isotopes, in 
                particular uranium and lead. ICP-MS is more suitable for this application than 
                the previously used thermal 
                ionization mass spectrometry, 
                as species with high ionization energy such 
                as osmium and tungsten can 
                be easily ionized. For high precision ratio work, multiple collector instruments 
                are normally used to reduce the effect noise on the calculated ratios. | 
            
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                 
                
                
             | 
         
     
    
        
            
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                 
                
                
             | 
            
                
                In the field of flow cytometry, 
                a new technique uses ICP-MS to replace the traditional fluorochromes. 
                Briefly, instead of labelling antibodies (or 
                other biological probes) with fluorochromes, each antibody is labelled with a 
                distinct combinations of lanthanides. 
                When the sample of interest is analysed by ICP-MS in a specialised flow 
                cytometer, each antibody can be identified and quantitated by virtue of a 
                distinct ICP "footprint". In theory, hundreds of different biological probes can 
                thus be analysed in an individual cell, at a rate of ca. 1,000 cells per second. 
                Because elements are easily distinguished in ICP-MS, the problem of compensation 
                in multiplex flow cytometry is effectively eliminated. | 
         
     
    
        
            | 
                
                In the pharmaceutical industry, ICP-MS is used for detecting inorganic 
                impurities in pharmaceuticals and 
                their ingredients. New and reduced maximum permitted exposure levels of heavy 
                metals form dietary supplements, introduced in USP (United States 
                Pharmacopeia) 
                <232>Elemental Impurities—Limits [1] and 
                USP <233> Elemental Impurities—Procedures,[2] will 
                increase the need for ICP-MS technology, where, previously, other analytic 
                methods have been sufficient. | 
            
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                 
                
                
             | 
         
     
    
    
        
            
                Quantification of proteins and biomolecules 
                
                In 2007, a new type of protein tagging reagents called metal-coded 
                affinity tags (MeCAT) 
                were introduced to label proteins quantitatively with metals, especially 
                lanthanides.[3] The 
                MeCAT labelling allows relative and absolute quantification of all kind of 
                proteins or other biomolecules like peptides. MeCAT comprises a site-specific 
                biomolecule tagging group with at least a strong chelate group which binds 
                metals. The MeCAT labelled proteins can be accurately quantified by ICP-MS down 
                to low attomol amount of analyte which is at least 2–3 orders of magnitude more 
                sensitive than other mass spectrometry based quantification methods. By 
                introducing several MeCAT labels to a biomolecule and further optimization of 
                LC-ICP-MS detection limits in the zeptomol range 
                are within the realm of possibility. | 
            
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                 
                
                
             | 
         
     
    
        
            
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                
                 
                
                
             | 
            
                Elemental analysis 
                    The ICP-MS allows determination of elements with atomic mass ranges 7 to 250 (Li to U), and sometimes higher. Some 
                    masses are prohibited such as 40 due to the abundance of argon in the sample. 
                    Other blocked regions may include mass 80 (due to the argon dimer), and mass 56 
                    (due to ArO), the latter of which greatly hinders Fe analysis unless the instrumentation is 
                    fitted with a reaction chamber. Such interferences can be reduced by using a 
                    high resolution ICP-MS (HR-ICP-MS) which uses two or more slits to constrict the 
                    beam and distinguish between nearby peaks. This comes at the cost of 
                    sensitivity. For example, distinguishing iron from argon requires a resolving 
                    power of about 10,000, which may reduce the iron sensitivity by around 99%. 
                
                    A single collector ICP-MS may use a multiplier in pulse counting mode to amplify 
                    very low signals, an attenuation grid or a multiplier in analogue mode to detect 
                    medium signals, and a Faraday cup/bucket to detect larger signals. A 
                    multi-collector ICP-MS may have more than one of any of these, normally Faraday 
                    buckets which are much less expensive. With this combination, a dynamic range of 
                    12 orders of magnitude, from 1 ppq to 100 ppm is possible. 
             | 
         
     
    
        
                                
        
                                ICP-OES Spectrometers Key 
        Features: 
                
         
        
                                
                
                
        
        
        
        
        
        
        
        
        
        
        
        
        
         
        
        
         
        
                
     
    
        
         
        Efficiency and Cost :  
    
        - 
            
                Powerful and cost effective alternative to 
                sequential ICP and FAA spectrometers. 
         
        - 
            
                Simultaneous spectrum capture in the 175-770 
                nm wavelength range - more than twice the sample throughput compared to 
                sequential instruments when more than 10 elements are analyzed. 
         
        - 
            
                Low operating cost - optical system requires 
                only 0.5 L/min purge gas while still offering superb analytical performance in 
                the UV spectral range. 
         
        - 
            
                Factory methods can be provided for standard 
                applications - "Plug and Analyze" without the need for method development. 
         
        - 
            
                Most accurate results and highest 
                sensitivity thanks to superior UV performance and constant resolution. 
         
        - 
            
                Faster out of the gate: less than 10 minutes 
                warmup time (previously over 30) with the new LDMOS generator. 
         
        - 
            
                Axial or radial plasma observation. 
         
     
    
        Power Source:  
    
        - 
            
                Next generation 
                plasma power, Laterally Difussed Metal Oxide Semiconductor (LDMOS) generator, 
                delivers up to 2000W. 
         
        - 
            
                
                
                High power gives great matrix compability at highest plasma loads previously 
                impossible. 
         
        - 
            
                
                
                Can actually analyze volatile organic samples such as gasoline at room 
                temperature ! 
         
        - 
            
                
                
 Sample can be analyzed in lower dilutions, for lower limit of detection. 
         
        - 
            
                
                
                Extremely robust and trouble free, with advanced design that is completely 
                short-cut-proof. 
         
        - 
            
                
                
                Air-cool system without external cooling, delivering low running costs. 
         
        - 
            
                
                
                Have no parts subject to wear - minimizing time and expanse. 
         
        - 
            
                
                
                Stablilize much more quickly in less than 10 minutes, ready to start most 
                measurements. 
         
     
    
        Excitation System:  
    
        - 
            
                Free-running generator with a frequency of 
                27.12 MHz, 0.7-2.0kW power output utilized. 
         
        - 
            
                For constant plasma power independent from 
                plasma load. 
         
        - 
            
                Perfectly suited for volatile organics and 
                high dissolved solids. 
         
        - 
            
                Includes automatic plasma ignition. 
         
        - 
            
                Energy and argon-saving standby mode. 
         
        - 
            
                Independently adjusts the resonance 
                frequency, holding the effective plasma power constant over a wide range of 
                sample loads - Characterized by its robustness, reliability and stability. 
         
     
    
        Sample Introduction System: 
     
    
        - 
            
                Extremely short fluid paths ensure short 
                measurement and flush times. 
         
        - 
            
                Thermally insulated illuminated compartment 
                provides plenty of room to accommodate a large variety of sample introduction 
                system. 
             
         
        - 
            
                Preadjusted setup makes for rapid startup, 
                without the need for comprehensive optimization. 
         
        - 
            
                Maintenance friendly sample introduction 
                system. 
         
     
    
        Plasma Interface: 
    
        - 
            
                Radial (SOP) or Axial (EOP) plasma 
                observation. 
         
        - 
            
                Radial observation 
                - Less sensitive to high matrix concentrations and the 
                influences from organic matrices. 
                 
                - Smaller observation volume for major elements high 
                precision determinations applications. 
         
        - 
            
                Axial observation: 
                 - Evaluates 
                the entire emission zone, facilitating the analysis of trace contents. 
                - OPI (Optical Plasma Interface), a bayonet connection 
                simplifies the use. 
                - Usually 10 times better detection limits without the 
                matrix interferences than other axial observation. 
                - Over 10 years proven. 
         
     
    
        Optical System: 
          
    
        - 
            
                Optimized  
                Rowland Circle Alignment(ORCA). 
         
        - 
            
                Covers the entire relevant wavelength range 
                from 175 to 777 nm. 
         
        - 
            
                Aluminum hollow section Design quarantees 
                high thermo stability measurement results. 
         
        - 
            
                Small optic volume guarantees short start up 
                times and a purge rate of less than l/min for measurements in the UV range (<200 
                nm), purged with Argon - optionally Nitrogen. 
         
        - 
            
                Line-rich spectra can be more easily 
                processed. 
         
        - 
            
                Delivers 8 picometers(pm) resolution in the 
                important range of 165nm-285nm. 
         
        - 
            
                Delivers 16 picometers(pm) resolution in 
                higher wavelengths. 
         
        - 
            
                A twin-interface version enables automatic 
                axial and radial observations. 
         
     
    
        Permanently Argon-filled optical chamber:
          
    
        - 
            
                UV gas purification system: Argon 
                circulation and cleaning system for sealed optical chamber. 
         
        - 
            
                Recirculating gas througth a small cleaning 
                catridge,good for 2 years of life. 
         
        - 
            
                Eliminates waste and 
                expense than conventional constant purge gas comsuming 
                 
                (typically more than S3000 per year that also may risks 
                system comtamination). 
         
        - 
            
                Helps achieve high stability and excellent 
                low wavelength performance. 
         
        - 
            
                Provides stable analytical results 
                immediately without purge delay and startup. 
         
     
    
        ICAL Logic System(Intelligent Calibration 
        Logic): 
    
        - 
            
                Continuously monitors the optical system and 
                normalizes the wavelength scale. 
         
        - 
            
                Calls for an 
                ICALization (measurement of the ICAL standard) if changes are noticed. 
         
        - 
            
                Carried out by the control logic if the 
                instrument is in automatic operation mode. 
         
        - 
            
                Methods can be easily 
                transferred among instruments, great for multiple systems of the same type 
                running. 
         
        - 
            
                Methods for numerous 
                and widely varying applications can be developed and offered as factory methods. 
         
     
    
        Readout System: 
    
        - 
            
                Linear CCD arrays. 
         
        - 
            
                Up to 8 orders of 
                magnitude large dynamic range detectors. 
         
        - 
            
                Possible to process concentrations from ppb 
                % range without blooming effects.
                
                
                
             
         
        - 
            
                High speed readout 
                system, parallelled digital signal processor units. 
         
        - 
            
                Recording and evaluation of the complete 
                spectrum in less than 3 seconds. 
         
     
    
        Cooling system: 
    
        - 
            
                No complicated, costely external colloing. 
         
        - 
            
                No water chiller 
                necessary, saving thousands while eliminating chiller breakdown risk. 
         
        - 
            
                No optic purge gas required, saves thousands 
                per year. 
         
        - 
            
                The only known 
                completely air-cooled ICP-OES analyzer, the unique innovative OPI-AIR interface. 
         
        - 
            
                Saves users from 
                having to buy, install, power, and maintain an elaborate expensive, and often short-lived external cooling system. 
         
     
    
        Software: 
    
        - 
            
                Offers theuser clear, structured and easy to 
                understand control of all instrument functions. 
             
         
        - 
            
                Consists of six main 
                modules for manual, automatic and transient measurements, spectra viewing, 
                method development and general configuration. 
         
     
    
        - 
            
                1-click routine operation and assistants for 
                the less frequently used options are available. 
         
        - 
            
                Even analytical 
                method creation is simplified vai intelligent assistants/wizards, default 
                settings, and predefined templates. 
         
        - 
            
                With dockable windows 
                and tool bars, the user interface can be customized to specific information 
                requirements and personal preferences. 
         
     
    
    
    
        - 
            
                All of the 
                measurement parameters, e.g., element and line selection can be edited and 
                changed as required. 
         
        - 
            
                Provide easiest 
                possible operation, offering manual, prompteed manual, or fully automatic 
                analysis, for personnel with minimal training. 
         
        - 
            
                Fully compatible with 
                US EPA, CLP as well as other standards and methods and includes an US FDA 21 CFR 
                Part 11 compliant user and data management. 
         
     
    
        Factory Calibrated Methods for wear metal in 
        oil: 
    
        - 
            
                Availiable with a 
                complete factory calibrated method set, ready with no need for initial method 
                development. 
         
        - 
            
                Swiftly and 
                accurately assesses component wear trends, detecting additives, wear elements 
                and contaminants. 
         
        - 
            
                Industrial-grade 
                durability, and surpringly affordable to purchase and operate. 
         
     
    
        Ergonomic: 
    
        - 
            
                Moderized ergonomic 
                design fits on any standard laboratory bench. 
         
        - 
            
                Polyurethane/aluminum 
                construction resistant to chemicals and corrosion. 
         
        - 
            
                Include thoughtful 
                layout features, as ultra-short fluid paths. 
         
        - 
            
                Easy, safe 
                accessibility to components from both sides and front. 
         
        - 
            
                For fast, convienient service or For fast, convienient service or 
                maintenance. 
         
     
    
                    
                More ICP-OES Spectrometers Key Features .....
                
             
    
        
                                
        
                                ICP-MS Spectrometers Key Features: 
        
     
    
        
                                
        
                                
                                
                
                
        
        
        
        
        
        
        
        
        
        
        
        
        
         
        
        
         
        
                
     
    
        - 
            
                Robust dry vacuum system and 
                zero-maintenance, free-running, generator. 
         
        - 
            
                Sample chamber can be accessed by sliding 
                the front cover to the side.
                
                
                
             
         
        - 
            
                Rapid exchange of the entire sample 
                introduction system or its individual components.
                
                
                
             
         
        - 
            
                Virtually every combination of nebulizer and 
                spray chamber is possible. 
         
        - 
            
                Easily developed by attaching 
                chromatographs, ultrasonic nebulizers or laser ablation systems. 
         
        - 
            
                Double-focusing sector field ICP-MS based on 
                a Mattauch-Herzog geometry with a newly developed ion optic and pioneering 
                detector technology.
                
                
                
             
         
        - 
            
                With a permanent ion beam going to the 
                detector. 
         
        - 
            
                ICAL system logic for continuous optimal 
                operation. 
         
        - 
            
                Capable of simultaneously measuring the 
                complete mass range used for inorganic mass spectrometry from 6Li to 238U. 
         
        - 
            
                Storage of the data for the complete mass 
                spectrum enables unprecedented flexibility regarding method development and 
                reprosessing. 
         
        - 
            
                Possible to later determine elements in a 
                sample even when the sample no longer exists. 
         
        - 
            
                First ICP-MS with simultaneously element 
                contents and isotope ratio analysis in the entire mass range from Li to U. 
         
        - 
            
                Rapid fingerprinting, real-time internal 
                standardization and measurement of transient signals, isotope ratios and isotope 
                dilution 
         
        - 
            
                Clearly better precision for development of 
                new application areas. 
         
        - 
            
                Leading in productivity and profitability 
                with high sample throughput rate. 
         
        - 
            
                Sets new standards for fully simultaneous 
                analysis, simple operation and highest reliability. 
         
        - 
            
                Tremendous ion detector readout that 
                achieves excellent detection limits, precision and accuracy. 
         
        - 
            
                Compatible with EPA, FDA, CLP and 21 CFR 
                Part 11 as well as other standards and guidelines. 
         
     
    
         
        More ICP-MS Spectrometers Key Features .....
         
     
    
        
                                
        
                                ICP-OES Spectrometers Application Report List: 
        
                
        *** Please login to request anapplication report
     
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
    
        
                                
                
        
                
                                 
        
                
        
                                
                ICP-MS Spectrometers Application Report List: 
        
                
        *** Please login to request anapplication report
     
    
    
    
    
    
    
    
    
    
    
         
     
                         |